Applications include Lab/Clean Room Pressurization, Filter Pressure, Fan Static Pressure.
Compatible pressure media: dry air and inert gases: nitrogen, helium, neon, argon, krypton, xenon.
High accuracy
Four ranges available
See Pressure Sensor Overview for Differential/Gauge and Absolute models
Included Tubing Adapter
Allows pressure connection to commonly available 1/8" I.D. tubing or 3.2 mm I.D. tubing. Consists of 8" length of 3/32 I.D. tubing and a 3/32" I.D. to 1/8" I.D. nylon adapter fitting.
Use any Pressure Scale
P450 Sensors can read in any unit of pressure (Pa, millibar, etc.); Simply convert the sensor's scaling values using the XR440 Pressure Converter for the XR440 or the XR5 Pressure Converter for the XR5.
Extension cable
The 18 inch leads on the P450 Pressure Sensor may be extended up to 100 FT (30 meters) by the user using three conductor 22 or 24 gage cable. In electrically noisy environments we recommend using shielded cable.
Compatible tubing
Commonly available 1/8" I.D. or 3.2 mm I.D. plastic tubing is compatible with the P450 using the included tubing adapter.
Tubing stocked at Pace
Clear 3/32" I.D. tubing connects directly to all P350, P400, P450 and P500 Sensors. Temperature range -40°C to 74°C (-40°F to 165°F). 5/32" (4 mm) O.D., minimum bend radius 3/8" (10 mm). Maximum pressure 43 PSI. Available in 25 FT (7.6 meter) rolls.
Part Nbr: T-332-25FT Price: $10.00 25 ft (7.6 meter) roll