P400-D-2-inch Pressure Sensor (±2 inches H2O)
Applications include Lab/Clean Room Pressurization, Filter Pressure, Fan Static Pressure.
Compatible pressure media: dry air and inert gases: nitrogen, helium, neon, argon, krypton, xenon.
Pressure range as low as ± 2 inch H2O
Miniature Size
High Accuracy
Low Power
Combined Linearity, Hysteresis & Repeatability: - ±0.25% of Span
Resolution: - See table below
Operating Temperature Range: -
Psi ranges: -25 to 90°C (-13 to 194°F)
±2 inch H2O range: -10 to 85°C (14 to 185°F)
Compensated Temperature Range: -
Psi ranges: -10 to 85°C (-4 to 185°F)
±2 inch H2O range: 0 to 60°C (32 to 140°F)
Total Error Band over compensated temperature range -
Psi ranges: 1% of Span
±2 inch H20 Range: 2% of Span
Supply Voltage - 4.75 to 5.25 Vdc
Power requirement: - 3.5 ma at 5 Vdc (supplied by Pace data logger)
Output signal -
0.5 to 4.5 Vdc at 5Vdc supply voltage.
Output is ratiometric to supply voltage.
Wiring: -
White: E terminal of Pace logger (Power)
Green: Analog Input of Pace Logger (Signal)
Black: "C" terminal of Pace logger (Ground)
Burst Pressure: - See table of available models (below).
Weight: - 0.3 oz. (8 grams)
Physical: - Epoxy encapsulated thermoplastic housing with three twisted leads, 18" length.
Pressure Ports: - Ports accept 3/32" I.D. tubing (or 2.4 mm I.D. tubing). Accepts 1/8" I.D. tubing (or 3.2 mm I.D. tubing) with included adapter kit (details below).
Suggested mounting methods: - Use cable tie or Velcro® straps to secure pressure hoses near sensor housing. Optional mounting base (see below)
Overall Size: - 0.9" x 0.6" x 0.75" (23mm x 15mm x 19mm)